Home Categories Send inquiry

Full-range ION Chromatograph

Technology advantage

1.Built-in circulating 3D constant temperature technology(CN 204259917U)
Temperature stability time is less than 30 mins, ensuring the accuracy and reliability of test data.
2.The world's leading full-range series of ion chromatographic columns(CN 105126936A、CN 104788603A)
High efficiency, large capacity of the columns for detecting ions of varied compositions.
3. Self-Regenerating Electrolytic Micro-membrane Suppressor(CN 102735792A)
High pressure resistance, small dead volume, highly responsive to signals.
4. Auto-range Conductivity Detector
It can directly detect the signal from ppb to ppm without adjusting the range; Only one conductivity detector can detect anions and cations.
5. Observatory intelligent workstation
Integrated control;With intelligent start-up, shutdown and maintenance functions;Compatible with a variety of instruments;Customized images.

Technical Parameter

Ion Chromatographic Pump
 
 
Type
 
High-pressure and
low-pulse two-piston
tandem advection pump
Maximum
Pressure
 
 
35MPa(PEEK)
Pressure
Display
Accuracy
 
 
 
≤0.1MPa
 
Flow Stability
 
(0.2-0.5)mL/min≤3%;
(0.5-1.0)mL/min≤2%;
>1.0mL/min≤2%;
Allowable Deviation of Flow
 
 
 
(0.2-0.5)mL/min , ±5%;
(0.5-1.0)mL/min, ±3%;
>1.0mL/min, ±2%;
Flow Range
 
 
0.001~9.999mL/min
Pressure Pulse
 
 
≤0.5%
Flow Precision  
≤0.1%
Flow Accuracy  
≤0.1%
 
 
 
Manual Sample Injector
Maximum Pressure
 
35MPa
External Diameter of the Tube 1/16"
Contact Material of the Rotor PEEK
Contact Material of  Medium PEEK/Ceramics
Column Heater

 

Operating Temperature Range 20~60℃(68~140℉)
Controlling Temperature Accuracy ±0.01℃
Allowable Deviation of Column Heater’s Temperature ±1℃
Temperature Stability ≤0.05℃/h
 Conduction
Detection System
Type Constant temperature
auto-range conductivity detector
Cell Volume ≤0.8μL
Detection Range 0~35000μS/cm
Detection Resolution ≤0.0020nS/cm
Output Voltage -6000+6000mv(adjustable)
Electronic Noise 0.02nS
Baseline Noise ≤0.001μS/cm
Baseline Drift ≤0.02μS
Operating Temperature Range Room temperature +5℃~60℃(41 ~140℉)
Controlling Temperature Accuracy ±0.01℃
Temperature Compensation 1.7 %/℃
Maximum Pressure 10.0MPa
Linear Range ≥103
Instrument Linearity ≥0.999
Quantitative Repeatability ≤1.0%
Qualitative Repeatability ≤0.5%
Minimum Detectable Concentration Cl-≤0.0005ug/mL;
Li+≤0.001ug/mL;
BrO3-≤0.001ug/mL;
Flow System Plastic Flow Path Made of PEEK materials
Six-way Valve PEEK material, pressure 5000psi;  Independent automatic collecting and flow
function.
Suppressor Type Self-Regenerating electrolytic micro-membrane suppressor
Maximum Pressure 6.0MPa
Dead Volume <50μL
Power Supply 150W
External Size(L*W*H*) 325*400*500
Net Weight(KG) 21
Gross Weight(KG) 28

Full-range ION Chromatograph Related Video: