Gas Chromatograph

Features

● The host adopts a 7-inch color touch screen with digital display of the flow (pressure) of the carrier gas/hydrogen/air channel.
● Lack of air alarm protection function; heating control protection function (when the oven door is opened, the oven fan motor and heating system are automatically turned off).
● The split/split ratio can be automatically controlled to save carrier gas.
● Configure autosampler to install positioning interface to match various specifications of autosampler.
● Multi-core 32-bit embedded hardware system to ensure reliable operation of the instrument.
● One-key start function, with 20 sets of sample test mode memory function.
● Logarithmic amplifier is adopted, the detection signal has no cut-off value, the peak shape is good, and the synchronous external trigger function can be expanded. The host and workstation can be started at the same time through external signals (autosampler, thermal analyzer, etc.).
● It has a perfect system self-check function and automatic fault identification function.
● With 8 external event expansion function interfaces, various function control valves can be selected, and work according to the time sequence set by yourself.
● RS232 communication port and LAM network port, as well as the configuration of the data acquisition card.

Column temperature box

● Content product: 22L
● Temperature control range: 5℃ ~ 400℃ at room temperature
● Temperature control accuracy: ±0.1℃
● Heating rate: 0.1 ~ 60℃ / min
● Program temperature rise order: 9
● Program heating repeatability: ≤ 2%
● Cooling way: open the door after
● Cooling speed: ≤10 mins (250℃ ~ 50℃)
Control software function :(GC112N only)
● Column temperature box control
● Detector control
● Injector control
● Map display

Sampler

● Temperature control range: 7℃ ~ 420℃ at room temperature
● Temperature control method: independent temperature control
● Carrier gas flow control mode: constant pressure
● Number of simultaneous installations: 3 at most
● Type of injection unit: filling column, shunt
● Split ratio: split ratio display
● Cylinder pressure range: 0 ~ 400kPa
● Cylinder pressure control accuracy: 0.1kPa
● Flow setting range:
H2 0 ~ 200ml / min
N2 0 ~ 150ml / min

Detector

● FID, TCD optional
● Temperature control: Max. 420℃
● Number of simultaneous installations: 2 at most
● Ignition function: automatic
Hydrogen flame ionization detector (FID)
Detection limit: ≤ 3×10 g/s (n-hexadecane)
Baseline noise: ≤ 5× 10-14A
Baseline drift: ≤ 6× 10-13A
Dynamic range: 107
RSD: 3% or less
● Thermal conductivity detector (TCD) :
Sensitivity: 5000mV•mL/mg (n-cetane)
Baseline noise: ≤ 0.05 mV
Baseline drift: ≤ 0.15mV / 30min
Dynamic range: 105
● Supply voltage: AC220V±22V, 50Hz±0.5Hz
● Power: 3000W