Gas Chromatograph Mass Spectrometer

product picture

Technical Specifications

Working Conditions
Power 220 V, 50 Hz
Temperature 15℃-35℃
Humidity 25%-80% RH
Specifications
Gas Chromatograph
Column oven temperature Room temperature + 10℃-400℃
Temperature stability ≤±0.03℃
Maximum heating rate 40℃/min
Maximum run time 999.99 min
10-segment programmable temperature control
Split/splitless inlet (3rd generation EPC)
Maximum temperature: 400℃
Electronic controlled pressure, flow rate and split ratio
Pressure range: 0-999 kPa
Flow range: 0-200 mL/min
Autosampler (optional)
Mass Spectrometer
Main Specifications
Mass range 1.5 - 1024.0 amu
Mass stability Better than 0.1 amu/48 h
Resolution Unit mass
Sensitivity DB-5MS 30m*0.25mm*0.25um fused silica capillary column or similar column.
EI source, full scan: (range 100-300 amu).
1 pg OFN  S/N≥100:1
Maximum scan rate 10,000 amu/s
Dynamic range 105
Ion source Electron impact ionization source (EI), standard.
Chemical ionization source (CI), optional.
Dual filaments Programmable switch
Maximum filament current 3 A
Emission current 10 – 350μA adjustable
Ionization energy 5 – 150eV adjustable
Ion source temperature 150 – 320℃ adjustable, individually controlled
Mass analyzer Quadrupole.
Full scan, selected ion monitoring (SIM) and acquisition.
At most 128 groups in SIM mode.
At most 128 ions in each group.
Detector electron multiplier + high-energy dynode back focusing assembly
GC-MS interface
Individually controlled through transmission cable, 150 – 320℃ adjustable
Vacuum system
Turbo molecular pump (250 L/s), mechanical pump (180 L/min)
Wide range compound cold cathode gauge
Data processing system
Hardware Computer (optional)
Printer Laser printer (optional)
Software MS3200RT Real-time data acquisition application and MS3200P data processing
application
Optional Accessories
DIP 100 liquid/solid direct injection probe assembly
Thermal desorption device
Dynamic headspace sampler
Purge-and-trap sample concentrator

Gas Chromatograph and Sample Introduction Unit

◆New industrial design, simple and generous, unique and humanized GC control panel. The design of the interface takes into account the protection of user errors.
◆ EPC gas control adopts the third-generation EPC control unit patented by our company, with pressure or flow control mode. The purge valve is electronically controlled to minimize sample diffusion and loss. The split/splitless injection mode can meet various application requirements. The automatic air-saving function effectively reduces operating costs. The original instant valve switch technology with no dead volume eliminates the long wait for stable pressure when the valve switch is opened in the splitless injection mode. This effectively improves the peak shape and repeatability of the retention time.
◆The optimized GC temperature control system improves the accuracy of the oven temperature control to ±0.03℃, which improves the repeatability of the analysis. The built-in installation positions of nitrogen, hydrogen and air gas circuits can be compatible with other chromatographic detectors according to user needs. The repeatability of the temperature program has been improved, resulting in sharper peaks of heavy compounds in the oil.
◆The unique CI reagent gas flow control module adopts feedback control, which automatically adjusts the reagent gas flow rate to the optimal level in proportion to the preset CI gas target ion, thereby saving the reagent gas while ensuring the high repeatability of the CI analysis.
◆Simple and practical direct liquid and solid injection probe options can perform rapid structural analysis of unknown compounds, providing a powerful tool for chemical synthesis applications. The unique detachable probe heater is easy to replace in case of damage or contamination. The highest temperature is 650°C.
◆Compatible with standard conventional GC columns.
◆Optional autosampler.
◆The software can be configured with a variety of optional peripheral accessories. Purge and trap concentrator, liquid autosampler, thermal desorption, headspace sampler, etc. can be easily set up, configured and controlled. Additional DO (digital output) port can be used for external device control
◆The innovative rotatable liquid autosampler can rotate 360° on the horizontal surface. The autosampler can be easily removed from the holder to simplify GC maintenance.
Powerful Software System:
Our user friendly software interface provides ease of use along with a powerful array of features for the advanced user. MS3200RT & MS3000P provide practical and accessible solutions for our user’s analytical needs.
MS3200RT Data Acquisition and Control Application
◆Chromatograms, mass spectra, parameters and instrument status are displayed simultaneously in a clean interface. Users can easily reference all relevant information during the analysis.
◆Available scanning modes include Scan, selected ion monitoring (SIM), or alternating Scan and SIM. Select scan mode based on desired analytical speed and quality.
◆All analysis parameters can be controlled through software, including carrier gas flow, pressure, column oven temperature, inlet temperature, etc. Automated GC-MS safe power down procedure can be initiated from the software.
◆Analysis method can be easily exported and imported.
◆Instrument status parameters are displayed in real time. Alarms are shown in noticeable colors. The automated low vacuum protection function protects fragile parts such as the filament, detector, etc.

MS3200RT Real-time high-speed data acquisition software

◆Total ion chromatogram and mass spectrum are displayed in the same interface to allow for easy comparison. Mass spectrum can be displayed as a processed bar graph or as raw data.
◆Snap spectrum transfer function by one click imports real-time files into the data processing software for qualitative and quantitative analysis.
◆The software offers a standard function menu for new users. Advanced users can use shortcut command keys for quicker access to features. Start, stop, and other actions can be performed by using accessible buttons on the main interface.
◆Both manual and automatic mass spectrum tuning are provided. Tuning conditions include resolution, sensitivity, abundance ratio, among others. These can be set according to analysis requirements. In manual tuning mode, effects of any changes in parameters on the mass signals can be observed. Manual tuning is suitable for both special application requirements and users with strong background in mass spectroscopy. Parameters and mass spectra are shown together for easy observation.
◆Software can perform a vacuum leak check function, which is essential for instrument maintenance.

Comprehensive Auto Tuning Interface

◆Instrument status can be monitored in the tuning interface so as to protect the instrument.
◆Switch between EI (electron ionization) and CI (chemical ionization) modes. Turn on/off the calibration compound.
◆Tuning reports can be quickly printed after tuning is complete.
◆Remote instrument diagnostic functions provide fast and professional technical assistance for your instrument anywhere in the world.
MS3200P Data Processing Application
◆All data processing methods are provided. The total ion chromatogram (TIC), mass spectrum, single ion chromatogram (MC), multiple ion chromatograms (MIC) are displayed on one screen for easy recognition and comparison of peak purity.
◆For qualitative analysis, the number of similar compounds displayed in the qualitative report can be set according to requirements. Report contents can be configured to obtain a simple qualitative report.
◆Quantitative functions include standard method, internal standard method, normalization method and corrected normalization method. MC, TIC, MIC can all be integrated upon and quantified.
◆The three-dimensional rendering function displays retention time, intensity, and mass number more intuitively in the same coordinate system.

Powerful Data Processing System

◆MS3200 MS software comes with professional petroleum data analysis tools for rapid analysis of petroleum chemical products for quality process control. Features include spectrum calculation, compound group screening and group composition export. The SNR calculation tool helps to evaluate the performance of instrument at any time. The spectrum addition and subtraction function is used to correct for the interferences caused by system background noise.
◆Data files are exported in CDF formats and can be imported by other software.
◆Other features include concise display layout, flexible qualitative approach of peaks, powerful batch processing capabilities and full quantitative methods
◆Standard spectra library provides manual single component query, and batch query. User defined libraries can be used for special applications.

Applications

The excellent performance of the GC-MS 3200 makes it suitable for applications in various fields including food safety, environmental safety, chemicals, among others.
●Detection of Melamine in Milk
●Analysis of Volatile Organic Compounds (VOC) in Drinking Water or Surface Water
●Detection of Plasticizer in Liquors
●Trace PAHs Detection
●Detection of Organochlorine Pesticide
●Rapid Semi-quantitative Analysis of Hydrocarbons
●Qualitative Analysis of Unknown Samples Using Direct Injection Probe

Flexible Configurations

◆The following configurations are applicable to water quality testing
(applicable to EPA Method 502.2)
Purge and trap analyzer + GC-MS 3200 + MS3200 software package + DB-624 (30 m × 0.25 mm × 1.4 μm) fused silica capillary column
Headspace sampler + GC-MS 3200 + MS3200 software package + DB-624 (30 m × 0.25 mm × 1.4 μm) fused silica capillary column
Applicable to quantitative detection of volatile organic compounds in surface water, drinking water and reservoir water.
◆Economical configuration applicable to ambient air quality monitoring
EW-3TD thermal desorption device + GC-MS 3200+ MS3200 software package + equivalent DB-5MS column (30 mx 0.25 mm x 0.25 μm) moderately polar column
Applicable to air quality testing in indoor environments and public spaces. Provides high sensitivity for TVOC and other common harmful gases.
◆Typical configuration applied to conventional laboratory analysis
Autosampler + GC-MS 3200 + MS3200 software package + DB-5MS (30m × 0.25mm × 0.25μm) fused silica capillary column
Suitable for qualitative and quantitative analysis of most organic compounds such as spices and perfumes, pesticides, batch sample analysis of PAHs.
◆Configuration applicable to quality control in a chemical synthesis process
DIP100 + GC-MS 3200+ MS3200 software package + DB-5MS (30m × 0.25mm × 0.25μm) fused silica capillary column
Applicable to rapid qualitative analysis of chemical synthesis intermediates and final products and quantitative analysis combined with GC sample introduction.
◆Mobile laboratory mounted in monitoring van
Analytical laboratory can be mounted to a mobile laboratory platform for rapid investigation of chemical contaminants in cases of food safety and environmental pollution emergencies.